Source and measure functions can be operated, and results displayed, simultaneously
Sources and reads DC voltage and current, resistance, temperature, contact, frequency and pressure
Accurate to 0.02%
Simulates and reads 3 RTDs and 8 thermocouples
Measures 2 or 3-wire RTDs
Compatible with 32 pressure modules from 2.5kPa to 70MPa
24V loop power supply
Provides 25% or 100% step output for DC current either manually or automatically
Large 2-line LCD with whie LED backlight
User selectable auto power off
Includes test leads, soft carrying case and batteries
Specifications: READ DC Voltage Range: 50mV, 500mV, 5V, 50V Accuracy: ±0.02% +10, ±0.02% +2, ±0.02% +5, ±0.03% +5 DC Current Range: 50mA Accuracy: ±0.02% +2 Resistance Range: 500Ω, 5kΩ Accuracy: ±0.02% +0.1Ω, ±0.02% +0.5Ω Frequency Range: 100Hz, 1kHz, 10kHz Accuracy: ±2% SOURCE DC Voltage Range: 100mV, 1V, 10V Accuracy: ±0.02% +10 DC Current Range: 20mA Accuracy: ±0.02% +10 Resistance Range: 400Ω, 4kΩ, 40kΩ Accuracy: ±0.02% +0.1Ω, ±0.05% +1Ω, ±0.1% +10Ω Frequency Range: 100Hz, 1kHz, 10kHz, 100kHz Accuracy: ±2% FUNCTION/READ RANGE/ACC'Y/SOURCE RANGE/ACC'Y Type R T/C/-40/1760°C/±1.5°C/-40/1760°C/±1.2°C Type S T/C/-20/1760°C/±1.5°C/-20/1760°C/±1.5°C Type K T/C/-200/950°C /±0.8°C/-200/1370°C /± 0.9°C Type E T/C/-200/500°C/±1.5°C/-200/1000°C/±0.5°C Type J T/C/-200/700°C/±0.7°C/-200/1200°C/± 0.7°C Type T T/C -200/400°C/±0.7°C/-200/400°C/± 0.6°C Type N T/C/-200/1000°C/±0.9°C /-200/1300°C/± 0.8°C Type B T/C/400/1800°C /±1.8°C/400/1800°C/± 1.1°C Pt100 RTD/-200/700°C/±0.8°C/-200/850°C/± 0.8°C Pt1000 RTD/-200/630°C/±0.4°C/-200/630°C/± 0.4°C Cu50/-50/150°C/±0.7°C/-50/150°C/±0.5°C